Document Page: First | Prev | Next | All | Image | This Release | Search
File: 102596_sep96_decls5_0011.txt
Subject: NAVGRAM LESSONS LEARNED F H F 1 JAN 91 TO 31 JAN 91
Unit: FLEET HOSP
Parent Organization: NAVCENTCOM
Box ID: BX300052
Folder Title: FLEET HOSPITAL WORKING PAPERS VARIOUS SOURCES CNA 2837
Document Number: 58
Folder SEQ #: 11
UNCLASSIFIED
(B) THE NEW JOHNSON & JOHNSON STERRAD GLOBAL DISCHARGE
UNIT, IS THE MOST PREFERRED. THE STERRAD IS ABOUT THE
SIZE OF AN EFFICIENCY APARTMENT REFRIGERATOR, USES A
110 VOLT STANDARD OUTLET. THE STERILIZATION CYCLE
TAKES APPROXIMATELY 50 MINUTES, AT A TEMPERATURE OF 100
DEGREES FAHRENHEIT, AND DOES NOT EMIT HARMFUL BY-
PRODUCTS, AS DOES ETO GAS STERILIZATION. THE STERRAD
SHOULD HAVE BEEN APPROVED BY THE FDA FOR SALES BY
JOHNSON & JOHNSON RECENTLY. WHICH GENERALLY SETS A
TREND FOR OTHER COMPANIES TO MARKET A SIMILAR PRODUCT,
SUCH AS AMSCO. THE STERRAD WOULD BE EXCELLENT FOR THE
TWO HOSPITAL SHIPS, WHICH ARE RESTRICTED BY THE COAST
GUARD TO USE ETO GAS STERILIZERS DUE TO SAFETY REASONS.
(9) INSTRUMENT CARE PRODUCTS SHOULD BE AVAILABLE FROM DAY
ONE AND THEN CONTINUOUSLY AVAILABLE VIA THE SUPPLY SYSTEM. THE
PRODUCTS INCLUDE:
(A) SONIC CLEANER/DETERGENT
(B) PHENOLIC DISINFECTANT/DETERGENT FOR
DECONTAMINATION AND TERMINAL CLEANING. THE IODINE
BASED PRODUCT DISINFECTANT-DETERGENT (NSN 6640-00-526-
1129) WESCODYNE, WHICH CAME WITH THE PLATFORM,M IS VERY
HARD ON INSTRUMENT SURFACES LEAVES STAINS, AND IS NOT
OPTIMUM AS A DECONTAMINATION AND CLEANING PRODUCT. A
PHENOLIC PRODUCT (SOME ARE CARRIED IN THE FEDERAL STOCK
SYSTEM) IS PREFERABLE AS AN ALL AROUND
DISINFECTANT/DETERGENT FOR ALL FLEET HOSPITAL USES,
INCLUDING COMMUNITY HEAD CLEANING, WARDS, CLINICS,
ICU-S, ANESTHESIA, ETC.
AS INSTRUMENT MILK (NSN
6515-02-247-8216, AN "L" ACQUISITION) FOR PRE-
STERILIZATION TREATMENT TO MAINTAIN PASSIVITY AND
SMOOTH JOINT MOVEMENT Or BOX LOCKS, ETC. (FROM A
DEPLOYED FLEET HOSPITAL STATUS, "L" ACQUISITION BECOMES
A MAJOR PROBLEM IN THE PROCUREMENT PROCESS)
(D) RUST REMOVER SOLUTION FOR REMOVING AND/OR
INHIBITING SPOTTING OR RUSTING, BAXTER HEALTH CARE HAS
A COMPATIBLE LINE OF PRODUCTS TO MEET THIS NEED.
(10) INSTALL CONSOLE SONIC CLEANERS IN EACH CSR ISO, FOR
CLEANING, RINSING AND DRYING CAPABILITIES. THE CONSOLE SONIC
CLEANERS NEED TO BE PLUMPED INTO A DISTILLED OR PURE WATER
SOURCE. RECOMMEND EITHER OF THE FOLLOWING SONIC CONSOLE CLEANERS
TO BE ADAPTED TO THE CSR MODULE:
(A) ZIMMER "DETRIX"
(B) CREST
(C) AMSCO
E. COMMENTS: COMMENTS ARE INCLUDED WITH DISCUSSION AND
RECOMMENDATION SECTIONS.
16. TANKER TRAILERS.
A. OBSERVATION: THE PUMPS ON THE 6000 GALLON TANKERS ARE
FOR USE WITH CLEAR LIQUIDS ONLY. SEWAGE CAN NOT BE PUMPED INTO
THE TANKERS USING THIS PUMP.
UNCLASSIFIED
Document Page: First | Prev | Next | All | Image | This Release | Search
Document 20 f:/Week-36/BX300052/FLEET HOSPITAL WORKING PAPERS VARIOUS SOURCES CNA 2837/navgram lessons learned f h f 1 jan 91 to 31 jan:1018961451347
Control Fields 17
File Room = sep96_declassified
File Cabinet = Week-36
Box ID = BX300052
Unit = FLEET HOSP
Parent Organization = NAVCENTCOM
Folder Title = FLEET HOSPITAL WORKING PAPERS VARIOUS SOURCES CNA 2837
Folder Seq # = 11
Subject = NAVGRAM LESSONS LEARNED F H F 1 JAN 91 TO 31 JAN
Document Seq # = 58
Document Date =
Scan Date =
Queued for Declassification = 01-JAN-1980
Short Term Referral = 01-JAN-1980
Long Term Referral = 01-JAN-1980
Permanent Referral = 01-JAN-1980
Non-Health Related Document = 01-JAN-1980
Declassified = 18-OCT-1996