File: 110196_aacbl_01.txtSUBJECT: FALSE POSITIVES USING M272 KITS AND WHIRL-PAKS PRIORITY PTTSZYUW RHIUFAA 0340700 RUEABOL ZNY P 030300Z FEB 91 FM USCENTAF//SGPB// TO AIG 8598//CC/SG// INFO RHDIAAA/HQ TAC LANGLEY AFB VA//SGPB// RUEABOL/HQ USAF BOLLING AFB DC//SGPA ZEN/USCENTAF//DOS// SUBJ: FALSE POSITIVES USING M272 KITS AND WHIRL-PAKS USCENTAF/SG # 91-FEB-011- 1. IT HAS BEEN IDENTIFIED BY SEVERAL.BASES IN THE AOR THAT FALSE~ POSITIVES FOR LEWISITE AND-CYANIDE ARE OBTAINED WHEN SAMPLING WATER WHICH HAS BEEN COLLECTED IN WHIRL-PAK THIOBAGS. IT IS ~ ~ SUSPECTED~THAT THE 10 MG OF SODIUM THIOSULFATE PRODUCES THE FALSE POSITIVE. EFFECTIVE IMMEDIATELY, ALL BIOENVIRONMENTAL ENGINEER1NG (BEE) OFFICES ARE TO REFRAIN FROM USING WHlRL-PAKS TO COLLECT WATER TO BE TESTED USING-THE M272 KIT. ANY CONTAINER~CAN, BE USE IN PLACE OF THE WHIRL-PAKS SINCE STERILITY IS NOT OF CONCERN 2.. (U) OUR THANKS TO THE836 TAC HOSP AND 1620 TAWP BEE OFFICES FOR PAGE 2 RHIUFAA0419 THIS HEADS-UP INFORMATION. DIRECT ANY QUESTIONS YOU MAY HAVE TO [(b)(2)] MINIMIZE CONSIDERED BT #0419 NNN SAVED AS FILE - C:\DAY034\XFAA0419.027 - 0340701Z